Models |
Digital Thermocouple Test Set & Calibrator Type : 7709 |
Digital Thermocouple Test Set & Calibrator Type : 7709C |
Digital Thermocouple Test Set & Calibrator Type : 7709EP |
Multifunction Process Calibrators Type : MC 50 & MC 75 |
Universal Pocket Calibrators Type : UC I |
Range of temperature |
-250ºC to 2500ºC |
- |
Ambient to 750ºC for 'J' typeAmbient to 1350ºC for 'K' typeAmbient to 1700ºC for 'S' typeAmbient to 1700ºC for 'R' typeAmbient to 400ºC for 'T' type |
(1). Background lit, graphic LCD Display and scaling function
(2). Digital, Menu driven value adjustment
(3). 22 keys and navigator
(4). Switchable menu languages
(5). Step-, ramp-, cycle-, Hold-, HART--protocol, filter
(6). Editable display units, real time clock,high speed call (%FS)
(7). Average Value, Nin/ Max, Offset, value set
(8). Stotage of 10 configuration files
(9). Transmitter supply 24 VDC / 22mA |
(1). Background lit, graphic LCD Display and scaling function
(2). Digital, Menu driven value adjustment
(3). Switchable menu languages
(4). Step-, ramp-, cycle-, Hold-, HART--protocol, filter
(5). Editable display units, real time clock, high speed call (%FS)
(6). Average Value, Nin/ Max, Offset, value set
(7). Stotage of 10 configuration files
(8). Serial USB PC interface (Type mini B) |
DC Volts measurement |
0 mV to ± 100.00 mV |
- |
- |
Calibrated potential source |
0 mV to ± 100 mV continuously variable through vernier and linear system. |
4 to 20 mA current transmitters. |
For the calibration of J, K, S, R, and T type of temperature indicators |
Accuracy DC Volt |
± 0.02% of f.s.d. ± 1 digit |
- |
± 0.3% of mV for T, J & K type thermocouples, ± 0.2% of mV for R & S type thermocouple |
Resolution |
0.1 ohm |
0.1 ohm |
0.1 ohm |
Accuracy |
± 0.1% |
± 0.1% |
± 0.3% of mV for T, J & K type thermocouples, ± 0.2% of mV for R & S type thermocouples |
Switch configuration |
Silver plated contact switches |
Silver plated contact switches |
Brass switches with silver plating |
Operating Temperature |
27ºC |
27ºC |
- |
Storage temperature |
40ºC to 60ºC |
40ºC to 60ºC |
- |
Relative Humidity |
45% RH + 10 % |
45% RH |
- |
Warm up period |
30 minutes |
30 minutes |
- |
Dimensions |
45(L) x 26(D) x 11(H) cm. |
47(L) x 29(D) x 14(H) cm |
48(L) x 35(B) x 13(H) cm. |
Weight |
Approx. 4 Kg. |
Approx. 6 Kg. |
7.5 Kg. |
Power Source |
230 V ± 10% AC, 50 Hz, perfectly stable and transients free AC Mains are required for stable digital read-out display. |
230 V ± 10% AC, 50 Hz |
230 V ± 10% AC, 50 Hz |