Impulse voltage of 100V-5kV, Eight channel Individual channel can be program and control Minimum inductance test value winding:10µH Up to 200MSa/s waveform sampling rate Maximum measuring speed: 6meas/sec Storage depth of 6k Bytes High bandwidth analog acquisition circuit High -fidelity corona extraction algorithm Four waveform comparision methods Automatic storage of instrument parameters Measurements on voltage, time and frequency Amplification,stretch and movement of the waveform for accurate display Multi sample average, average processing of 32 standard waveform Use demagnetized impulse to ensure the conformity of tested waveform 20 groups of instrument files storage and retrieval Screen information storage in USB disk System firmware up gradation facility through USB-disk Foot control interface for easy measurements Handler interface to realize on-line operation 65k color 7" TFT wide display screen RS232C, USB Device and LAN interface to realize remote control
SME1201-1 | SME1205-1 | SME1205-4 | SME1205-8 | SME1210-1 | ||
No. of channels | 1 | 1 | 4 | 8 | 1 | |
Output Impulse Voltage | 30V-1200V | 100V 5000V | 500V 0 V | |||
5V steps | 10V Steps | 20V Steps | ||||
Voltage Accuracy | ± (5% of set value + 5V) | ± (5% of set value + 15V) | ± (5% of set value + 25V) | |||
Read back Accuracy | ± (5% actual value + 5V) | ± (5% actual value +15V) | ± (5% actual value + 25V) | |||
Inductance Test Range | >=1µH | >=10µH | >=20µH | |||
Impulse Energy | Max : 0.02 Joule | >Max : 0.25 Joule | >Max : 0.5 Joule | |||
Test Speed | 6 meas/ second | 6 meas/ second | 3 meas/ second(when 10kVimpulse voltage is output) | |||
Pulses Applied | Max : 32 | |||||
Input Impedance | 5M Ohm | |||||
Waveform Acquisition | Sampling Rate Max 200 MSa/s, 8 levels adjustable Resolution 8 Bits Memory Depth 6k Bytes Average : 1 to 32 | |||||
Comparison Mode | Area size comparison,Differential area comparison,Corona discharge comparison, Differential phase comparison | |||||
Waveform Measurement | Voltage, Time, Frequency | |||||
Trigger Mode | Internal/ Manual (foot)/ External/ BUS | |||||
Detection Output | PASS/ FAIL display /LED/Alarm | |||||
Measurement Statistics | Statistics for Measurement Results | |||||
Memory | 20 groups of standard waveform data & instrument setup can be stored in internal non-volatile memoryUSB flash memory can be used as external memory | |||||
Interface | HANDLER, RS232C, USB Device, USB Host, LAN | |||||
Gernal Information | ||||||
Display | 800X480 Dots, 65k color TFT, Waveform Display Range : 600x256 | |||||
Safety and electromagnetic compatibility | IEC61010-1:2001, IEC61326-2-1:2005 | |||||
Supply | 220V ± 10 %, 50/60Hz ± 5% | |||||
Power consumption | =<200VA/td> | |||||
Operating Conditions | 0°C 40°C, 5% RH | |||||
Weight | 7.2 kg (Approx.) | |||||
Standard Accessories | Foot Switch, High-voltage test clip leads, User Manua |